Technical artifact simulating "motor axonopathy".

J M Gilchrist1, E Dosremedios, G Johnson

  • 1Department of Neurology, Rhode Island Hospital, Brown University, Providence.

Muscle & Nerve
|April 1, 1992
PubMed
Summary

Technical artifact in nerve conduction studies can mimic motor axonopathy. A partial electrode transection caused falsely low compound motor amplitudes, resolved by replacing the electrode.