Geometric misalignment and calibration in cone-beam tomography
Related Experiment Video
Updated: Aug 5, 2026

Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
Lorenz von Smekal1, Marc Kachelriess, Elizaveta Stepina
1Institut für Medizinische Physik, Universität Erlangen-Nürnberg, Henkesstr 91, D-91052 Erlangen, Germany. Lorenz.von.Smekal@imp.uni-erlangen.de
Abstract:
We present a new high-precision method for the geometric calibration in cone-beam computed tomography. It is based on a Fourier analysis of the projection-orbit data, recorded with a flat-panel area detector, of individual point-like objects. For circular scan trajectories the complete set of misalignment parameters which determine the deviation of the detector alignment from the ideal scan geometry are obtained from explicit analytic expressions. To derive these expressions we show how to disentangle the problems of calculating misalignment parameters and point coordinates. The calculation of the coordinates of the point objects inside the scanned volume, in units of the distance from the focal spot to the center of rotation, is then possible analytically likewise. We simulate point-projection data on a misaligned detector with various amounts of randomness added to mimic measurement uncertainties. This data is then employed in our calibration to validate the method by comparing the resulting misalignment parameters and point coordinates to the known true ones. We also present our implementation and results for the geometric calibration of micro-CT systems. The effectiveness of the corresponding misalignment correction in reducing image artifacts is exemplified by reconstructed micro-CT images.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Errors in Taping
Adjusting a Traverse