Related Experiment Video
Updated: Aug 19, 2026

Assessment of Neuromuscular Function Using Percutaneous Electrical Nerve Stimulation
Published on: September 13, 2015
Is the stimulation frequency of the repetitive nerve stimulation test that you choose appropriate?
1Department of Neurology, College of Medicine, National Cheng-Kung University, Tainan, Taiwan.
Abstract:
The repetitive nerve stimulation test (RNST) has been a useful method in the diagnosis of myasthenia gravis (MG). In clinical practice, a short train of repetitive stimulation is usually given at 3 Hz. Although it was documented that lower stimulation frequencies could offer a greater sensitivity, no study has been done to testify the most sensitive stimulation frequency for RNST. To find out an optimal stimulation frequency, we performed RNST at 0.5, 1, 3, 5, 7, 10, 15 and 20 Hz in 15 MG patients and 5 healthy subjects. The results showed that the decremental response was most often seen at 7 Hz rather than at 3 Hz. To augment the sensitivity in the diagnosis of MG, RNST should be performed stimulation not only at 3Hz but also at other frequencies, preferably 7 Hz.