Reproducibility and stability of neuromagnetic source imaging in primary somatosensory cortex
Michael Schaefer1, Nina Nöennig, Anke Karl
1Human Cortical Physiology Section, National Institute of Neurological Disorders and Stroke, National Institutes of Health, Bethesda, Maryland, USA. schaefem@ninds.nih.gov
Abstract:
The present study investigated the test-retest reliability of magnetoencephalography (MEG) source localization of somatosensory evoked fields (SEFs) over an extended time period. Five healthy subjects were stimulated pneumatically at the first and fifth digit in two sessions spaced several months apart. At each location 400 stimuli were presented. The validation of the results was performed by overlay of the dipole localizations into the individual anatomic structure of the subjects' cortex by the use of magnetic resonance images (MRIs). The source localizations of the SEF component were found to be highly reproducible. The mean standard deviation of the dipole locations of the first digit was 1.55 mm in the x-, 1.55 mm in the y- and 3.49 mm in the z-direction. The mean standard deviation of the fifth digit was 3.69 mm in the x-, 4.27 mm in the y- and 6.60 mm in the z-direction. These results support the use of MEG recordings combined with MRI as an adequate method to define the organization of the human primary somatosensory cortex and provide a useful approach to the rapid detection of neuroplasticity.
More Related Videos
09:57Electromagnetic Source Imaging in Presurgical Evaluation of Children with Drug-Resistant Epilepsy
Published on: September 20, 2024
14:34Surgical Protocol for a Large, Resealable Cranial Window Enabling Longitudinal, Multi-Modal Electrophysiology Recordings of Mouse Default Mode Network
Published on: May 29, 2026