Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Do metallic endoprostheses increase radiation dose associated with automatic tube-current modulation in
Stefania M R Rizzo1, Mannudeep K Kalra, Michael M Maher
1Division of Abdominal Imaging and Intervention, Department of Radiology, Massachusetts General Hospital and Harvard Medical School, White 270-E, 55 Fruit St., Boston, MA 02114, USA.
Objective:
Our objective was to assess the effect of orthopedic metallic prostheses on radiation dose associated with MDCT using z-axis automatic tube-current modulation and a fixed tube current in a phantom and patient study.
Conclusion:
Z-modulation is associated with a 34.1% increase in the mean tube current-time product and no change in the extent of streak artifacts in patients with a metallic prosthesis, compared with patients without a prosthesis. However, compared with the fixed-tube-current technique, z-modulation is associated with a 28.9% decrease in the mean tube current-time product.