Small tip angle three-dimensional tailored radiofrequency slab-select pulse for reduced B1 inhomogeneity at 3 T
Suwit Saekho1, Fernando E Boada, Douglas C Noll
1Department of Bioengineering, University of Pittsburgh, Pittsburgh, Pennsylvania 15213, USA.
Magnetic Resonance in Medicine
|January 29, 2005
Abstract:
This work presents a small tip-angle 3D tailored RF slab-select pulse for reducing the B1 field inhomogeneity at 3T. The compensated slice profile was determined from a B1 inhomogeneity map. SNR improvement and degree of artifact reduction were evaluated in a NiCl2 doped phantom and human brains. The technique was found to reduce inhomogeneities as large as 30% of the peak image magnitude in all three spatial directions in the brain using a standard head coil.