Related Experiment Video
Updated: Aug 19, 2026

Registered Bioimaging of Nanomaterials for Diagnostic and Therapeutic Monitoring
Published on: December 9, 2010
SQUID-detected MRI at 132 microT with T1-weighted contrast established at 10 microT--300 mT
Seung Kyun Lee1, Michael Mössle, Whittier Myers
1Department of Physics, 366 LeConte Hall, University of California, Berkeley, CA 94720, USA. Isk@socrates.berkeley.edu
Abstract:
T(1)-weighted contrast MRI with prepolarization was detected with a superconducting quantum interference device (SQUID). A spin evolution period in a variable field between prepolarization and detection enabled the measurement of T(1) in fields between 1.7 microT and 300 mT; T(1) dispersion curves of agarose gel samples over five decades in frequency were obtained. SQUID detection at 5.6 kHz drastically reduces the field homogeneity requirements compared to conventional field-cycling methods using Faraday coil detection. This allows T(1) dispersion measurements to be easily combined with MRI, so that T(1) in a wide range of fields can be used for tissue contrast. Images of gel phantoms with T(1)-weighted contrast at four different fields between 10 microT and 300 mT demonstrated dramatic contrast enhancement in low fields. A modified inversion recovery technique further enhanced the contrast by selectively suppressing the signal contribution for a specific value of the low-field T(1).