Intracardiac voltage gradients during transthoracic defibrillation: implications for postshock myocardial injury
James T Niemann1, Robert G Walker, John P Rosborough
1Harbor-UCLA Medical Center, Department of Emergency Medicine, 1000 West Carson Street, Box 21, Torrance, CA 90509, USA. jniemann@emedharbor.edu
Unlabelled:
In-vitro studies indicate that the electric-field intensity, or voltage gradient (VG), generated by a defibrillation shock is a determinant of defibrillation success as well as potential shock-induced cardiac injury. It is not clear how common descriptors of shock dose, e.g., joules (J), relate to VGs.
Objectives:
To assess the relationships between shock energy descriptors and VG.
Methods:
One monophasic and three biphasic waveforms were compared using transthoracic shocks and standard electrodes in five swine. VG measurements via intracavitary multielectrode-tipped catheters were compared with delivered energy and peak current. Shock variables were recorded at native transthoracic impedance and an adjusted impedance approximating that of typical humans.
Results:
For shocks at the same energy setting, peak current and VGs varied widely among the four defibrillators. At simulated human impedance (75 Omega), shocks at each device's maximum energy setting produced similar VGs among the three biphasic defibrillators, despite different delivered energies. For both native impedance (35 Omega) and at 75 Omega, VG correlated with peak current (r = 0.81 and 0.77, respectively) but not energy setting (r = 0.61, 0.52) or delivered energy (r = 0.58, 0.56). At the maximum energy setting of each device, maximum recorded VGs for both monophasic (33 V/cm) and biphasic (24 V/cm) defibrillators were less than those reported to cause myocardial injury (>60 to 80 V/cm).
Conclusions:
Energy descriptors correlate poorly to actual shock intensities. When compared with reported VG thresholds of myocardial injury, this study suggests that risk of injury from critically strong VGs is low for all of these defibrillators and equivalent among tested biphasic waveforms.
More Related Videos
09:47A Rat Model of Ventricular Fibrillation and Resuscitation by Conventional Closed-chest Technique
Published on: April 26, 2015
08:19Transthoracic Echocardiography to Assess Post-Resuscitation Left Ventricular Dysfunction After Acute Myocardial Infarction and Cardiac Arrest in Pigs
Published on: July 12, 2022
Related Concept Videos
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Electrophysiology of Normal Cardiac Rhythm
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...