Objective:
To determine whether tympanostomy tube (TT) inner diameter or shaft length impacts the rate of mucoid plug clearance.
Study Design And Setting:
Ex vivo model. Silicone TTs with different inner-diameters (ID) and shaft-length (SL) pairings (1.14 mm ID x 12 mm SL versus 1.14 mm ID x 1 mm SL; 1.14 mm ID x 4.8 mm SL versus 1.32 mm ID x 4.8 mm SL) were plugged with middle-ear mucus (n = 15 per group) and placed in a model ear chamber. Ofloxacin otic solution was instilled into the chamber to cover the plugged TT, and the time to clearance of each plug was recorded.
Results:
TTs with larger IDs ( P = 0.019) and greater SLs ( P = 0.033) cleared plugs more rapidly. However, the difference in the percentage of tubes that unplugged was not significant ( P = 0.151).
Conclusions:
Rate of ex vivo TT plug clearance may be altered by changing TT ID and SL.
Related Concept Videos
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Anatomy of the Ear
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask and...