Related Experiment Video
Updated: Aug 19, 2026

Deep Proteome Profiling by Isobaric Labeling, Extensive Liquid Chromatography, Mass Spectrometry, and Software-assisted Quantification
Published on: November 15, 2017
Use of lysozyme as a standard for evaluating the effectiveness of a proteomics process
Leah S Riter1, Barry D Hodge, Karen M Gooding
1Lilly Research Laboratories, Eli Lilly and Company, Indianapolis, Indiana 46285, USA.
Abstract:
Automated sequencing of unknowns in bottom-up proteomics makes the data produced susceptible to process control errors, which can be propagated into mistakes in analyte identification. Inclusion of an unintrusive internal standard, such as lysozyme, allows monitoring all phases of the proteomics process including sample preparation, enzymatic digestion, HPLC, mass spectrometry, and database searching. By using this internal standard, digestion issues including rearrangements, semi-tryptic fragments, and modifications were monitored. In addition, control of the HPLC process including column performance was achieved. The use of the lysozyme standard allowed easy optimization of mass spectral conditions including data dependent and collision induced dissociation settings. The use of this internal standard in a study of differential protein expression in rat serum samples is presented.