Accuracy in Dental Medicine, A New Way to Measure Trueness and Precision
Published on: April 29, 2014
A comparison of 18 different x-ray detectors currently used in dentistry
Allan G Farman1, Taeko T Farman
1Department of Surgical and Hospital Dentistry, The University of Louisville, Louisville, KY 40292, USA. agfarm01@louisville.edu
Purpose:
There has been a proliferation of available dental x-ray detectors over the recent past. The purpose of this short technical report is to provide a basic comparison of spatial resolution, contrast perceptibility, and relative exposure latitudes of 18 current dental x-ray detectors, including solid-state systems (CCD and CMOS), photostimulable phosphors, and analog film.
Methods:
Spatial resolution was measured using a 0.025 mm Pb phantom test grid with a measurement range from 1.5 to 20 lp/mm. For contrast perceptibility, a 7-mm thick aluminum perceptibility test device with wells of 0.1-0.9 mm depth at 0.1 mm intervals and 1 defect of 1.5 mm was used. The relative exposure latitude was determined by expert consensus using clear discrimination of the enamel-dentin junction as the lower limit and pixel blooming or unacceptable levels of cervical burnout as the upper limit.
Results:
The highest spatial resolution was found with Kodak InSight film, RVG-ui (CCD), and RVG 6000 (CMOS) detectors, each of which achieved 20 lp/mm, followed by the Planmeca Dixi2 v3 at > or =16 lp/mm. Contrast resolution was at least to 0.2 mm through 7 mm aluminum for all 18 detectors, with the best results being found for the Visualix HDI, RVG-ui, RVG 5000, and RVG 6000 detectors and the Schick CDR wired and wireless systems. The greatest exposure ranges were found with photostimulable phosphors and the Kodak RVG 6000 and RVG 5000 detectors.
Conclusions:
Most current x-ray detectors generally perform well in terms of spatial and contrast resolutions, and in terms of exposure latitude. These findings were independent of the modality applied.
More Related Videos
Related Concept Videos
X-ray Imaging
Determination of Crystal Structures
X-ray Diffraction of Biological Samples
According to Bragg's law, when X-rays strike the sample positioned on a stage, the rays are scattered by the electron clouds around the sample atoms. The X-ray diffraction or scattering is caused by constructive interference of the X-ray waves that reflect off the internal crystal...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT