Effects of skin pigmentation on pulse oximeter accuracy at low saturation
Philip E Bickler1, John R Feiner, John W Severinghaus
1Department of Anesthesia and Perioperative Care, University of California at San Francisco, 513 Parnassus Avenue, San Francisco, CA 94143, USA. bicklerp@anesthesia.ucsf.edu
Background:
It is uncertain whether skin pigmentation affects pulse oximeter accuracy at low HbO2 saturation.
Methods:
The accuracy of finger pulse oximeters during stable, plateau levels of arterial oxygen saturation (Sao2) between 60 and 100% were evaluated in 11 subjects with darkly pigmented skin and in 10 with light skin pigmentation. Oximeters tested were the Nellcor N-595 with the OxiMax-A probe (Nellcor Inc., Pleasanton, CA), the Novametrix 513 (Novametrix Inc., Wallingford, CT), and the Nonin Onyx (Nonin Inc., Plymouth, MN). Semisupine subjects breathed air-nitrogen-carbon dioxide mixtures through a mouthpiece. A computer used end-tidal oxygen and carbon dioxide concentrations determined by mass spectrometry to estimate breath-by-breath Sao2, from which an operator adjusted inspired gas to rapidly achieve 2- to 3-min stable plateaus of desaturation. Comparisons of oxygen saturation measured by pulse oximetry (Spo2) with Sao2 (by Radiometer OSM3) were used in a multivariate model to determine the interrelation between saturation, skin pigmentation, and oximeter bias (Spo2 - Sao2).
Results:
At 60-70% Sao2, Spo2 (mean of three oximeters) overestimated Sao2 (bias +/- SD) by 3.56 +/- 2.45% (n = 29) in darkly pigmented subjects, compared with 0.37 +/- 3.20% (n = 58) in lightly pigmented subjects (P < 0.0001). The SD of bias was not greater with dark than light skin. The dark-light skin differences at 60-70% Sao2 were 2.35% (Nonin), 3.38% (Novametrix), and 4.30% (Nellcor). Skin pigment-related differences were significant with Nonin below 70% Sao2, with Novametrix below 90%, and with Nellcor at all ranges. Pigment-related bias increased approximately in proportion to desaturation.
Conclusions:
The three tested pulse oximeters overestimated arterial oxygen saturation during hypoxia in dark-skinned individuals.
More Related Videos
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring pulse
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Hypoxia
Types of Hypoxia
There are four primary types of hypoxia, each resulting from a different cause:
1. Anemic hypoxia: This type occurs due to insufficient oxygen delivery caused by a lack of red blood cells (RBCs) or RBCs with abnormal or...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...