Performance of orientation sensors for use with a functional electrical stimulation mobility system
Related Experiment Video
Updated: Aug 18, 2026

Home-Based Monitor for Gait and Activity Analysis
Published on: August 8, 2019
Scott Simcox1, Simon Parker, Glen M Davis
1Rehabilitation Research Centre, The University of Sydney, Australia. ssimcox@fhs.usyd.edu.au
Abstract:
The purpose of this study was to verify the performance of recently developed body-worn sensor packs against 3D motion analysis of trunk and lower-limb movements. Five sensor packs, each consisting of rate gyroscope and two 2-D accelerometers controlled by a microprocessor were attached to the trunk, thighs, and shanks of an able bodied subject. A 6-camera motion analysis system (MAS) recorded multiple trials of sit-to-stand movements and normal walking. Time domain signals from each sensor pack were significantly correlated (r = 0.90-0.99;p < 0.05) with a root mean square errors of less than 5 degrees when compared against the same limb angle measurements calculated by the MAS. These data demonstrate that these external sensor packs are accurate devices for measuring trunk and lower-limb sagittal plane orientation in real-time.