Insights from the sharp end of intravenous medication errors: implications for infusion pump technology
M Husch1, C Sullivan, D Rooney
1Patient Safety Team, Northwestern Memorial Hospital, Feinberg School of Medicine, Northwestern University, 676 North St Claire, Chicago, IL 60611, USA. mhusch@nmh.org
Intravenous medication errors are frequent with IV pumps, but current smart pump technology alone is insufficient for significant patient safety improvements. Interfacing smart pumps with other systems is crucial for enhancing medication safety.
Area of Science:
- Patient Safety
- Medical Device Technology
- Health Informatics
Background:
- Intravenous (IV) medication errors are a significant concern in hospitals, with numerous FDA-reported incidents involving IV pumps leading to patient harm and deaths.
- Understanding the types, frequency, and severity of these errors is critical for developing effective safety measures.
Purpose of the Study:
- To identify the types, frequency, and severity of medication errors associated with IV pump administration.
- To assess the potential of smart pump technology, without system integration, to prevent these errors.
Main Methods:
- A point prevalence study prospectively observed IV pump medication administration.
- Researchers compared pump settings (medication, dose, rate) against the medical record.
- Preventability by smart pumps was retrospectively evaluated based on current technology capabilities.
Main Results:
- Of 426 observed IV medications, 285 (66.9%) had administration errors.
- A total of 389 errors were documented; most were not anticipated to cause patient harm (NCC MERP category C).
- Only one error could have been prevented by standalone smart pump technology.
Conclusions:
- Medication errors with IV pumps are common, diverse, and potentially harmful.
- Smart pumps are essential for safe medication systems but require integration with systems like EMR and CPOE for meaningful safety gains.
- Future research should explore technology's effectiveness in preventing errors and investigate new error types introduced by technology.
More Related Videos
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
IV Infusion to Oral Dosing: Conversion Methods
Parenteral Drug Delivery Systems: Injectables, Implants, and Infusion Devices