Related Experiment Video
Updated: Aug 18, 2026

Metal-silicate Partitioning at High Pressure and Temperature: Experimental Methods and a Protocol to Suppress Highly Siderophile Element Inclusions
Published on: June 13, 2015
[Determination of non-rare earth impurities including Cu, Fe, Ca and Ni in samarium oxide by ICP-AES]
Abstract:
The direct determination of non-rare earth trace impurities including Cu, Fe, Ca and Ni in high purity Samarium oxide by means of ICP-AES is presented in this paper. By using orthogonality experimentation, we studied main factors of influential determination and selected compromise condition of simultaneous determination of Cu, Fe, Ca and Ni. Influence of matrix concentration was discussed. We used the method of matrix match and subtraction background. For the sample solution containing 5mg/mL of Sm2O3 the determination limits were 0. 0003% for Cu, 0.003% for Ni, 0.001% for Ca and Fe, respectively. The relative standard deviation was in the range of 2.6-9.2%. The recovery ratio of these elements was in the range of 90-110%. The method is simple and rapid. It was fit to productive control analysis and determination of production.
More Related Videos
Related Concept Videos
Atomic Emission Spectroscopy: Lab
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Precipitation and Co-precipitation
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...