Purpose:
To theoretically calculate refractive changes in eyes with reversed intraocular lenses (IOLs).
Setting:
Nara Medical University, Nara, Japan.
Methods:
With the ray-tracing method, the refractive changes in a theoretical eye model with a reversed IOL in the capsular bag were calculated. Data for 3 models, anterior biconvex (MA60AC, Alcon), posterior biconvex (MA60BM, Alcon), and convex plano (UV2565T, Menicon), were provided by the manufacturer and used in the analysis.
Results:
Using the assumption that the axial length varied from 22 to 27 mm, the anterior-biconvex IOLs elicited small refractive changes when reversed. The posterior-biconvex IOLs elicited a significant myopic shift (up to 0.90 diopter [D]) when reversed. The convex-plano IOLs elicited a significant hyperopic shift (up to 2.01 D) when reversed.
Conclusions:
Anterior-biconvex IOLs with similar anterior and posterior radii of curvature will not cause large diopter surprises when reversed. High-power posterior-convex IOLs will cause significant myopic diopter surprises when reversed. Both these changes shift toward myopic if the postoperative anterior chamber depth (ACD) becomes shallow. Medium- to high-power convex-plano IOLs will cause a large hyperopic shift when reversed. This hyperopic shift decreases if the postoperative ACD becomes shallower.