The Stealth Station Image Guidance System may interfere with pulse oximetry
Johannes H van Oostrom1, Michael E Mahla, Dietrich Gravenstein
1Department of Anesthesiology, University of Florida College of Medicine, P.O. Box 100254, Gainesville, Florida 32610-0254, USA. hans@anest.ufl.edu
Purpose:
Interference on pulse oximetry can come from many sources. We found an additional source of interference from the Stealth Station. This article gives an overview of sources of pulse oximeter interference so that clinicians can better prevent them. Technical features: This article discusses the infrared disturbances caused by the Stealth Station. The Stealth Station is a frameless stereotactic positioning system that utilizes a three dimensional location system to measure the position of the patient and the surgical tools, and to relate those positions to previously recorded imaging. To understand the disturbance caused by the Stealth Station, we discuss its operation and that of pulse oximeter monitors. Pulse oximeter interference can come from volume artifacts, electrical and light noise, and can be caused by issues related to the patient. Because the passive Stealth Station contains a strong infrared light source, interference caused by light is a likely reason for the interference we noted. Pulse oximeters rely on the time-variant light signal modulated by arterial volume variations in the finger. Although relatively immune to static light sources, pulse oximeters are extremely sensitive to time-varying light sources. The light emitted by the passive Stealth Station is time-varying at 4 Hz and this is causing the pulse oximeter to provide invalid results. Shielding can generally be used to stop the light from the Stealth Station from being picked up by the pulse oximeter sensor.
Conclusion:
Infrared light interference can be very common, but is easily dealt with if one is aware of it.
More Related Videos
06:57Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Special considerations while measuring pulse
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...