Related Experiment Video
Updated: Aug 18, 2026

A Uniform Shear Assay for Human Platelet and Cell Surface Receptors via Cone-plate Viscometry
Published on: June 5, 2019
Platelet activation can occur by shear stress alone in the PFA-100 platelet analyser
1Department of Haematology, Prince of Wales Hospital, Level 4 Campus Centre, Barker St, Randwick, 2031, Sydney, Australia. dunkley@sesahs.nsw.gov.au
Abstract:
The PFA-100 platelet function analyser simulates high-shear platelet function; however, the relative importance of the high shear rates alone has not been examined in this system. Normal blood was analysed by the PFA-100 using modified cartridges that contained a blank membrane lacking the usual platelet agonists (collagen/epinephrine or collagen/ADP) thus isolating the effects of shear stress alone. We found evidence of platelet activation with significant increases in platelet-leucocyte aggregates. There was no significant change in surface CD62P nor PAC-1 expression. A degree of platelet activation occurs in the PFA-100 due to shear stress alone.