Defibrillation threshold testing: is it really necessary at the time of implantable cardioverter-defibrillator
Andrea M Russo1, William Sauer, Edward P Gerstenfeld
1University of Pennsylvania Health System, Philadelphia, Pennsylvania 19104, USA. andrea.russo@uphs.upenn.edu
Objectives:
The purpose of this study was to (1) determine how often implantable cardioverter-defibrillator (ICD) system modifications were needed to obtain an adequate safety margin for defibrillation, (2) identify how often and for what indications defibrillation threshold (DFT) testing was not performed, and (3) identify factors predicting the need for modification.
Background:
Ventricular fibrillation (VF) typically is induced at the time of ICD insertion. Although DFT testing often is minimized, a safety margin of 10 J has been utilized as a standard of care. However, current devices offer technology such as biphasic waveforms and high outputs, and the need for testing has been questioned.
Methods:
We reviewed the records of the last 1,139 patients undergoing initial ICD placement, generator replacement, or revision.
Results:
Seventy-one patients (6.2%) were identified as having an unacceptably high DFT (<10 J safety margin) requiring intervention, and some required >1 modification. Use of a high-output device alone was not enough to obtain an adequate DFT in 48% (34/71) of patients who required modifications (3% of the total population). No arrhythmia inductions were performed in 54 patients (4.7%) because of well-defined clinical conditions. Patients who required system modification had a lower ejection fraction, were younger, were less likely to have coronary artery disease, were more likely to be undergoing upgrade/generator replacement, and were more likely to be taking amiodarone. Long-term mortality was not different between the group of patients who required modification compared with those who did not (17% vs 20%, P = NS).
Conclusions:
Routine VF induction and documentation of effective defibrillation still remains a reasonable part of ICD placement because an inadequate safety margin may occur in >6% of patients. The incidence of patients who were inappropriate for testing based on well-defined clinical conditions is small (<5%) in this unselected large series. Although some clinical features may predict the need for system modification, additional studies are needed to better define "acceptable efficacy" of ICDs in preventing sudden death prior to altering these standards in selected patients.
Related Concept Videos
Cardiopulmonary Resuscitation III: AED Use
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias V: Evaluating Dysrhythmias
Dysrhythmias VI: Management of Dysrhythmias
Cardiopulmonary Resuscitation IV: Pharmacological Management
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...