Related Experiment Videos
Clinical reliability of insertion gain measurements with assistive listening devices
B McPherson1, L Hickson, A Baumfield
1Department of Speech and Hearing, University of Queensland, Brisbane, Australia.
Scandinavian Audiology
|January 1, 1992
Abstract:
Real ear insertion gain was measured in 14 persons fitted with a variety of common assistive listening devices. Measurements were made in a quiet, but not soundproofed, clinical environment. The reliability of such insertion gain measures was studied, both for short-term and long-term test-retest conditions. The results indicated that insertion gain measurement of assistive listening devices can be performed as a routine clinical procedure. Intrasubject standard deviations were comparable in most cases to those clinically obtained using hearing aids with custom ear moulds.