Related Experiment Video
Updated: Aug 18, 2026

In Vitro Evaluation of The Effects Of Er,Cr:YSGG and Diode Lasers Used on Titanium Cylinder
Published on: June 6, 2025
In vitro study of the variable square pulse Er:YAG laser cutting efficacy for apicectomy
Josko Grgurević1, Lovro Grgurević, Ivana Miletić
1Department of Oral Surgery, School of Dental Medicine, Gundulićeva 5, 10 000 Zagreb, Croatia. josko@sfzg.hr
Background And Objectives:
Variable square pulse (VSP) Er:YAG laser should be quicker than older Er:YAG lasers. The objectives were: (1) comparison of VSP laser and mechanical handpiece efficacy for apicectomy and (2) determination of optimal pulse width/energy/frequency combination.
Study Design/Materials And Methods:
Sixty extracted, single-rooted mature human teeth with round apical parts were instrumented, root filled, cleaned, and divided into four groups. Apical 2 mm of each root were apicectomized with mechanical handpiece and Er:YAG laser with three different settings (LaserA = 200 mJ/300 microseconds/ 8 Hz; LaserB = 200 mJ/100 microseconds/8 Hz; LaserC = 380 mJ/100 microseconds/20 Hz). Timing results were statistically compared.
Results:
LaserC was the most efficient setting. Differences between groups were significant except between LaserC-Mechanical and LaserA-LaserC (P < 0.05).
Conclusions:
VSP Er:YAG laser used for apicectomy is slower by a factor of 7-31 than mechanical handpiece, but treatment outcome is acceptable. Optimal settings for apicectomy with VSP laser are 380 mJ/100 microseconds/20 Hz.