Related Experiment Video
Updated: Aug 18, 2026

Plant Sample Preparation for Nucleoside/Nucleotide Content Measurement with An HPLC-MS/MS
Published on: February 24, 2021
Development of a solid-phase extraction/gas chromatographic-mass spectrometric method for quantification of succinic
Ake Stenholm1, Ingrid Drevin, Michael Lundgren
1Process Analytical Chemistry, Amersham Biosciences, Björkgatan 30, S-75184 Uppsala, Sweden. ake.stenholm@amersham.com
Abstract:
A solid-phase extraction (SPE)/gas chromatographic-mass spectrometric (GC-MS) method was developed for analysing residual succinic acid in nucleoside derivatives to be used in oligonucleotide synthesis. Use of a SPE protocol, enabled most of the derivatives to be trapped, thereby creating eluates enriched in succinic acid. GC-MS was used to quantify the amount of residual succinic acid in four different nucleoside preparations, with succinate concentrations varying from 0.18 to 0.24% (w/w). The within-day repeatability of the method was found to be 1.25% RSD. A linear relationship was observed between the amount of succinic acid in the sample and the GC-MS peak area, with a correlation coefficient of 0.9997 in the concentration interval 0.05-2.5% (w/w). Recoveries were measured by the addition of internal standards to working solutions and varied between 99.8 and 102.6%.