Neonatal resuscitation 2: an evaluation of manual ventilation devices and face masks
C P F O'Donnell1, P G Davis, R Lau
1Royal Women's Hospital Melbourne, Victoria 3053, Australia. colm.odonnell@rwh.org.au
Background:
The key to successful neonatal resuscitation is effective ventilation. Little evidence exists to guide clinicians in their choice of manual ventilation device or face mask. The expiratory tidal volume measured at the mask (V(TE(mask))) is a good estimate of the tidal volume delivered during simulated neonatal resuscitation.
Aim:
To compare the efficacy of (a) the Laerdal infant resuscitator and the Neopuff infant resuscitator, used with (b) round and anatomically shaped masks in a model of neonatal resuscitation.
Methods:
Thirty four participants gave positive pressure ventilation to a mannequin at specified pressures with each of the four device-mask combinations. Flow, inspiratory tidal volume at the face mask (V(TI(mask))), V(TE(mask)), and airway pressure were recorded. Leakage from the mask was calculated from V(TI(mask)) and V(TE(mask)).
Results:
A total of 10,780 inflations were recorded and analysed. Peak inspiratory pressure targets were achieved equally with the Laerdal and Neopuff resuscitators. Positive end expiratory pressure was delivered with the Neopuff but not the Laerdal device. Despite similar peak pressures, V(TE(mask)) varied widely. Mask leakage was large for each combination of device and mask. There were no differences between the masks.
Conclusion:
During face mask ventilation of a neonatal resuscitation mannequin, there are large leaks around the face mask. Airway pressure is a poor proxy for volume delivered during positive pressure ventilation through a mask.
More Related Videos
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)