Pressures generated in vitro during Stabident intraosseous injections
J M Whitworth1, R A M Ramlee, J G Meechan
1School of Dental Sciences, Newcastle University, Newcastle upon Tyne, UK. j.m.whitworth@ncl.ac.uk
International Endodontic Journal
|May 7, 2005
Summary
The Stabident intraosseous injection technique does not pose a significant risk of dangerous pressure build-up in local anesthetic cartridges when used as directed. Studies show pressures generated are below the threshold for anesthetic cartridge failure, even with rapid injections.
Area of Science:
- Dental Anesthesiology
- Biomechanical Engineering
Background:
- Intraosseous anesthesia techniques aim for rapid onset and profound anesthesia.
- The Stabident system is an intraosseous injection device requiring specialized perforation.
- Concerns exist regarding potential high-pressure risks and anesthetic cartridge failure with intraosseous methods.
Purpose of the Study:
- To evaluate the hypothesis that the Stabident intraosseous injection technique generates high pressures.
- To assess the risk of anesthetic cartridge failure associated with the Stabident system.
- To quantify intra-cartridge pressures during Stabident injections under various conditions.
Main Methods:
- A modified dental syringe was used to measure local anesthetic cartridge pressures.
- Injections were performed into air and sheep mandible models (young/old) simulating Stabident perforations.
- Both slow (approx. 15s) and rapid (<10s) injection rates of 2% Xylocaine with 1:80,000 adrenaline were tested.
- Pressure data were recorded at 1-second intervals, with statistical analysis including one-way ANOVA.
Main Results:
- Injection pressures were significantly higher when injecting into tissue compared to air (P<0.001).
- Rapid injections generated significantly higher intra-cartridge pressures than slow injections (P<0.05).
- Higher pressures were observed in younger, smaller sheep mandibles compared to older, larger ones. The maximum recorded pressure was 3.31 MPa, below the glass cartridge fracture threshold.
Conclusions:
- Stabident intraosseous injections, when performed according to manufacturer guidelines, do not present a serious risk of dangerous pressure build-up.
- The pressures generated are insufficient to cause local anesthetic cartridge failure.
- The technique is deemed safe concerning anesthetic cartridge integrity.