Pacing with capture verification in candidates for resynchronisation therapy: a feasibility study.
Mauro Biffi1, Giuseppe Boriani, Matteo Bertini
1Institute of Cardiology, University of Bologna, Via Massarenti 9, 40138 Bologna, Italy. mbiffi@orsola-malpighi.med.unibo.it
Summary
Autocapture during biventricular pacing (BIV) is feasible and safe, verifying capture on the second ventricular channel. This approach may enhance cardiac resynchronisation therapy (CRT) device longevity and reduce healthcare costs.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Devices
Background:
- Cardiac resynchronisation therapy (CRT) devices use multiple output channels, increasing energy consumption and reducing longevity.
- Autocapture technology has previously improved pacemaker longevity and safety in right ventricular pacing.
- Investigating Autocapture in biventricular pacing (BIV) aims to reduce the energy demands of CRT.
Purpose of the Study:
- To assess the feasibility and safety of applying Autocapture during biventricular pacing (BIV).
- To determine if Autocapture can reduce the energy cost of CRT.
- To evaluate the accuracy of evoked response (ER) detection during BIV pacing.
Main Methods:
- An acute study during BIV device implantation tested ER detection on the ventricular channel paced second.
- The study compared ER amplitudes during BIV pacing versus single-ventricle pacing (VVI mode).
- Long-term Autocapture performance was monitored by tracking RV thresholds over 20 months.
Main Results:
- Evoked response (ER) amplitudes in both ventricles were not significantly reduced during BIV pacing compared to VVI pacing.
- Autocapture successfully verified capture on the RV channel during BIV pacing, even with LV depolarization.
- Long-term monitoring showed consistent Autocapture performance without sensing LV depolarization as an ER on the RV channel.
Conclusions:
- Capture verification using Autocapture on the secondarily paced ventricular channel during BIV pacing is feasible and safe.
- This method has the potential to increase the longevity of CRT devices.
- Implementing Autocapture in BIV pacing may lead to reduced healthcare costs for heart failure patients.