Related Experiment Videos
Hydroxyapatite coating on carbon composite hip implants in dogs
G L Maistrelli1, N Mahomed, D Garbuz
1Toronto East General and Orthopaedic Hospital, Ontario, Canada.
Abstract:
In 33 dogs we implanted femoral stems made either of carbon composite, some coated with hydroxyapatite, or of titanium alloy with a porous coating. Osseointegration was greater in the hydroxyapatite-coated than in the un-coated stems (p less than 0.001). Push-out tests, at an average of 7.2 months after implantation, showed a six-fold increase in interface shear strength and a twelve-fold increase in shear stiffness in the hydroxyapatite-coated group compared with noncoated implants. The highest shear-strength values were found in the porous-coated titanium alloy stems, around which there was also the most resorptive bone remodelling.