Do we measure bilirubin correctly anno 2005?
Joke J Apperloo1, Fedde van der Graaf, Volkher Scharnhorst
1Clinical Chemistry Laboratory, Máxima Medical Centre, Veldhoven, The Netherlands. j.apperloo@mmc.nl
Clinical Chemistry and Laboratory Medicine
|May 19, 2005
Summary
Discrepancies in total bilirubin measurements between liquid and dry chemistry analyzers were observed in human serum. These differences, linked to bilirubin variants, highlight that reference material agreement doesn't ensure accuracy in patient samples.
Area of Science:
- Clinical Chemistry
- Analytical Chemistry
- Biochemistry
Background:
- Accurate total bilirubin determination is crucial for diagnosing and monitoring liver conditions.
- Interlaboratory agreement for bilirubin measurements is often sought using standard reference materials.
- Liquid and dry chemistry analyzers may exhibit different performance characteristics for bilirubin assays.
Purpose of the Study:
- To investigate the discrepancies in total bilirubin measurements between liquid and dry chemistry analyzers.
- To evaluate the performance of standard reference material SRM916 in assessing inter-analyzer agreement.
- To identify the factors contributing to observed discrepancies in human serum samples.
Main Methods:
- Comparison of total bilirubin concentrations measured by liquid and dry chemistry analyzers.
- Utilizing the Jendrassik-Grof method as a reference.
- Analysis of standard reference material SRM916 and human adult serum samples.
- Investigating the impact of conjugated and delta-bilirubin on assay results.
Main Results:
- A 30% discrepancy was observed between liquid and dry chemistry analyzers for human serum total bilirubin.
- Liquid analyzers showed a 20% overestimation, while dry analyzers showed a 10% underestimation relative to the reference method.
- SRM916 demonstrated excellent agreement across analyzers and near 100% recovery against the reference method.
- Discrepancies in human serum were attributed to the presence of conjugated or delta-bilirubin.
Conclusions:
- Inter-analyzer agreement with reference materials does not guarantee accurate bilirubin measurements in human serum.
- The presence of specific bilirubin fractions (conjugated, delta) significantly impacts assay performance.
- Careful consideration of analyzer type and sample matrix is necessary for reliable bilirubin quantification.