X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Potential of dose reduction after marker placement with full-field digital mammography
Christopher C Riedl1, Silvia Jaromi, Daniel Floery
1Memorial Sloan-Kettering Cancer Center, New York, NY 10021, USA. riedlc@mskcc.org
Objectives:
The purpose of our study was to assess the potential for radiation dose reduction in digital postinterventional digital mammograms after marker placement.
Materials And Methods:
One hundred consecutive cases of marker placement (hook-wire localization or postbiopsy clip marker placement), with 200 full-field digital baseline mammograms (craniocaudal and mediolateral), were included in this prospective trial. For the postinterventional digital mammograms, the milliampere seconds were reduced either by 50% or by 75%. Dose-reduced images were evaluated for sufficient image quality to verify the position of the marker.
Results:
In 193 of 200 cases (96.5%), image quality was sufficient to verify the correct position of the marker. One (1%) case with insufficient image quality occurred in the 50% dose-reduction group and 6 (6%) in the 75% dose-reduction group (P = 0.06).
Conclusion:
Our results indicate that under evaluation of each individual case, a dose reduction of 50% to 75% can be recommended in postinterventional digital mammograms.