Review of Swanson trapezium implants
1Microsurgery Research Centre, St Vincent's Hospital, Melbourne, Victoria, Australia.
The Australian and New Zealand Journal of Surgery
|June 1, 1992
Summary
The Swanson trapezium silicone prosthesis shows early problems, including dislocation and pain, making it unsuitable for osteoarthritis thumb treatment despite high patient satisfaction.
Area of Science:
- Orthopedic Surgery
- Biomaterials Science
Background:
- Osteoarthritis of the thumb is a common condition causing pain and functional limitation.
- Silicone trapezium prostheses are used to treat thumb osteoarthritis.
- The Swanson trapezium prosthesis is one such implant.
Purpose of the Study:
- To retrospectively review the outcomes of Swanson trapezium silicone prostheses.
- To evaluate the efficacy and complications of this implant in treating thumb osteoarthritis.
Main Methods:
- Retrospective review of sixteen Swanson trapezium silicone prostheses.
- Average follow-up of 3 years.
- Analysis of patient satisfaction, functional compromise, pain, implant stability, and radiographic findings.
Main Results:
- High patient satisfaction reported.
- 14 out of 16 thumbs had residual functional compromise and pain.
- 3 implants dislocated, 4 were subluxed.
- 5 thumbs showed intra-osseous cysts or bony erosion.
- 1 patient required implant removal.
Conclusions:
- The Swanson trapezium silicone prosthesis is associated with significant early problems.
- Complications include implant instability, pain, and bone erosion.
- The procedure is deemed unsuitable for treating osteoarthritic thumbs.