Chest radiography: a comparison of image quality and effective dose using four digital systems
A Pascoal1, C P Lawinski, A Mackenzie
1ME&P, King's College London, London, UK. ana.pascoal@kcl.ac.uk
Abstract:
Image quality (IQ) and effective dose for chest radiography was compared for four digital imaging systems that used three different detector technologies: a-Si/TFT flat-panel detector (FPD), scanning-slot/charge-coupled device (CCD) and photostimuable phosphor (PSP). On each system a phantom was exposed at 125 kV(p) for automatic exposure control (AEC) and 2, 4 and 8 Gy receptor dose using identical geometrical conditions. All images were scored as softcopy images by three observers. The effective dose was calculated for each exposure condition. For AEC, superior IQ was observed for the GE FPD compared with all the other systems, which showed similar IQ performance. For all systems the entrance surface dose associated with AEC was within the European recommendations but variations in the effective dose were observed between the four systems. For identical receptor dose levels superior IQ was observed with the FPDs. Thorascan was noted for its low effective dose and Agfa CR was associated with the highest effective dose. FPD systems showed a better overall performance, followed by the CCD and PSP systems.
Related Concept Videos
X-ray Imaging
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT