Peter Leman1, Shaun Greene, Kim Whelan
1Emergency Department, St Thomas' Hospital, London, UK. peter.leman@health.wa.gov.au
Objective:
To compare the novel Boussignac valve continuous positive airways pressure (CPAP) delivery mask and a standard closed-circuit Drager CF800 CPAP system in the management of acute pulmonary oedema (APO) patients.
Methods:
This was a randomized controlled trial whereby patients presenting to the ED with APO and who met the study criteria received either CPAP via the Boussignac valve system or from a standard Drager CF800. Baseline physiological and arterial gas data were recorded and repeated at 30 and 60 min after CPAP commenced. The primary outcome was mean change in pCO2 at 60 min between the two systems.
Results:
There were 39 evaluable patients (19 Boussignac, 20 Drager). The mean change in pCO2 at 60 min compared to baseline was similar in the two groups (Boussignac 0.9 kPa vs. Drager 1.2 kPa, mean difference -0.3; 95% CI -1.0-0.5, P=0.45). In addition, there were no significant differences at 60 min in regards to respiratory rate decrease, Boussignac 17.3/min versus Drager 19.6/min (mean difference 1.3; 95% CI -3.3-5.8, P=0.58) or peripheral SaO2 increase, Boussignac 10.7% versus Drager 14.6% (mean difference -3.9; 95% CI -9.9-2.1, P=0.19). There was no significant difference in disposition from the ED or the complication rate.
Conclusions:
The Boussignac valve system may be an effective lightweight disposable method of delivering CPAP to patients with APO. It appears to perform as effectively as much larger, more expensive and less transportable equipment.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...