Related Experiment Video
Updated: Oct 11, 2026

Troubleshooting FoCUS Image Acquisition: Patient Positioning, Transducer Manipulation, and Image Optimization
Published on: March 3, 2023
Artificial Intelligence Echocardiographic Interpretation in the Emergency Department: Agreement, Interpretation Yield
Vijay Manivel1,2, Michael Hession3,4, Behzad Mirmiran1
1Emergency Department, Nepean Hospital, Sydney, Australia.
Objective:
To assess the agreement between artificial intelligence (AI) interpretation and expert emergency physician sonologist review of echocardiograms performed in the Emergency Department (ED).
Methods:
Retrospective multicentre agreement study in four Australian EDs. A total of 195 echocardiograms (2016-2023) underwent three-stage adjudicated human expert review. Human reports were compared with us2.ai reports. Agreement across 10 validated parameters was reported as percentage exact match with 95% confidence intervals and kappa under two pre-specified tiers: Guideline Compliant (GC; ASE/EACVI multi-view criteria) and Pragmatic (PR; any available measurement). Reference standard reliability was quantified.
Results:
Pooled agreement was 76% at both tiers (GC n = 875; PR n = 1323); AI returned an interpretation in 46% of evaluations at GC and 70% at PR. Four structural parameters reached 78%-87%: right atrial size 87% (κ = 0.62), right ventricular (RV) contractility 83% (κ = 0.42), left ventricular (LV) size 83% (κ = 0.52) and RV size 78% (κ = 0.37). Binary LV systolic agreement was 79% GC and 83% PR. Expert reviewers agreed on 85% of interpretations. Pericardial effusion, aortic regurgitation and mitral stenosis, present in 7%-20% of patients, are not reported in the current us2.ai algorithm.
Conclusion:
AI agreed with expert review in about three-quarters of evaluations but provided a guideline-compliant interpretation in less than half; a pragmatic approach raised yield to 70% at the same agreement. The AI withheld classifications that experts could reach; its validated scope covers only part of what ED echocardiography must address, and reports did not mark unassessed parameters. This study reports agreement, not accuracy. For now, human expert review of AI reports is required.
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Acute Coronary Syndrome III: Diagnostic Studies