A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Electrogram configuration and detection of supraventricular tachycardias by a morphology discrimination algorithm in
L Lüthje1, D Vollmann, M Rosenfeld
1Kardiologie und Pneumologie, Georg-August-Universität Göttingen, Robert-Koch-Strasse 40, D-37099 Göttingen, Germany. larsluethje@med.uni-goettingen.de
Background:
Inappropriate ICD therapy for supraventricular tachycardia (SVT) remains a significant problem. A morphology-based algorithm (Wavelet) compares baseline and tachycardia electrograms (EGM). For this analysis different EGM sources can be programmed. This study evaluates the performance of Wavelet using two different EGM configurations (SVC-Can and RV-Can) for the detection of exercise-induced SVT.
Methods:
Patients with a Medtronic model 7230 single chamber ICD and a dual coil lead were included. For each EGM source (SVC-Can or RV-Can), a baseline EGM template was acquired and the morphology similarity to this template (match percentage) was evaluated for 10-15 beats at different heart rates during exercise testing. The lower VT detection limit was programmed to 600 ms (therapies off).
Results:
A total of 28 patients (66.9 +/- 4.7 years, 93% men) and 5,824 intracardiac QRS complexes were analyzed. With the RV-Can source, a consistently high similarity to the baseline EGM template was observed (< or =100 bpm: 90.90 +/- 0.56%; >100 bpm: 90.24 +/- 0.55%, P > 0.05). In contrast, SVC-Can was associated with a lower match percentage at baseline and a significant decrease at higher heart rates (< or =100 bpm: 77.91 +/- 2.65%; >100 bpm: 59.05 +/- 5.65%, P < 0.005). Accordingly, the specificity for appropriate detection of exercise-induced SVT was higher with RV-Can (21/21 episodes) than with SVC-Can (8/18 episodes, specificity 100% vs 44%; P < 0.0001).
Conclusion:
The RV-Can configuration appears to be superior to SVC-Can as EGM source for appropriate SVT detection with the Wavelet algorithm.
More Related Videos
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Dysrhythmias III: Characteristics of Dysrhythmias
Electrophysiology of Normal Cardiac Rhythm