[Physical property of emission data in simultaneous emission/transmission acquisition: evaluation for clinical
Fumitoshi Nakamura1, Toshihiko Kanno, Tatsuo Torizuka
1Positron Medical Center, Hamamatsu Medical Center.
Purpose:
Although simultaneous emission transmission acquisition (SET) is the ideal method for positron emission tomography (PET) because the same patient position can be used for both emission and transmission scans, the quality of PET images is subject to interference by the cross-contamination of each data. In recent years, segmented attenuation correction of transmission data has made it possible to exclude the contamination from emission data. In the present study using a phantom, the physical property of emission data in SET acquisition was evaluated in comparison with that in separate scan.
Methods:
We measured scatter fraction, scatter projection, % random, and noise equivalent count rate in the sinogram. Next, we determined the acquisition time of the SET scan to obtain the same coefficient of variation as the separate scan in reconstructed images.
Results:
In the evaluation of the strength of the line source, the physical property of emission data in the SET scan seemed inferior to that in separate scan, and SET scanning was not effective enough to reduce the acquisition time of PET examinations.
Conclusion:
SET scanning has the advantage that the same patient position can be used for emission and transmission data; however, further studies may be necessary to apply it to clinical examinations.
More Related Videos
14:19A Basic Positron Emission Tomography System Constructed to Locate a Radioactive Source in a Bi-dimensional Space
Published on: February 1, 2016
08:01Rapid Scan Electron Paramagnetic Resonance Opens New Avenues for Imaging Physiologically Important Parameters In Vivo
Published on: September 26, 2016
Related Concept Videos
Atomic Emission Spectroscopy: Overview
Atomic Emission Spectroscopy: Instrumentation
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Atomic Emission Spectroscopy: Lab
Absorption of Radiation
Transmission Electron Microscopy