Quality control of slope-intercept measurements of glomerular filtration rate using single-sample estimates
John S Fleming1, Linda Persaud, Maureen A Zivanovic
1Department of Medical Physics and Bioengineering, and Nuclear Medicine, Southampton University Hospitals NHS Trust, UK. John.Fleming@suht.swest.nhs.uk
Objectives:
Measurement of glomerular filtration rate (GFR) using the slope-intercept technique determines the plasma clearance curve by fitting a straight line to the logarithm of sample count rate. When two samples are used there is no check on the validity of curve fitting. GFR may also be estimated from single-sample concentrations. This study describes a method of quality control for the two-sample technique using the agreement between the one-sample and two-sample estimates.
Methods:
GFR measurements using Tc-DTPA were performed on 225 adults and 100 children using two samples taken between 2 h and 4 h post-injection. The two-sample values obtained using the British Nuclear Medicine Guidelines slope-intercept technique were compared to one-sample estimates obtained using a new general equation. Equations describing the variation of GFR error with GFR value were defined. These were used to determine action levels giving the limits of expected agreement between slope-intercept and single-sample values. The use of these action levels for quality control was demonstrated in a further 120 GFR measurements.
Results:
The variation of single-sample error estimate with GFR depended both on the time of sample and body surface area. For specific sample groups, the error variation with GFR could be approximated using a truncated quadratic equation. Four studies were identified as failing quality control in the dataset used to define the error equations. Two studies failed in the test dataset.
Conclusions:
One-sample equations give reliable estimates of GFR, which may be used for quality control of slope-intercept GFR assessment.
Related Concept Videos
Renal Clearance
Renal clearance refers to the volume of plasma cleared of a specific substance, such as creatinine, per unit of time. To measure clearance, urine samples are collected over a 24-hour period during each bladder voiding, followed by a single blood sample at the...
Renal Drug Excretion: Glomerular Filtration
Drugs gain access to the kidney via the renal artery, which progressively branches off into afferent arterioles.
Determination of Renal Drug Clearance: Graphical and Midpoint Methods
The graphical method involves plotting the rate of drug excretion in urine against the plasma drug concentration. By analyzing the graph, the clearance can be calculated and obtained. Drugs rapidly excreted by the kidneys exhibit a...
Drug Dosing in Renal Diseases: Measurement of Glomerular Filtration Rate
Drug Dosing in Renal Diseases: Measurement of Serum Creatinine Concentration and Clearance
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration