Propagation of measurement errors in glomerular filtration rate determination: a comparison of slope-intercept,
Jennifer L Holness1, John S Fleming2, Maxwell T Chirehwa3,4
1Department of Medical Imaging and Clinical Oncology, Nuclear Medicine Division, Stellenbosch University.
Background:
Measurement errors occurring during glomerular filtration rate (GFR) studies propagate to an error in the calculated GFR. Previous work has modelled measurement errors for slope-intercept (SI-GFR), single-sample (SS-GFR) and slope-only (SO-GFR) methods. In this study, we have extended these models. The primary aims were to (i) compare measurement errors in two-sample SI-GFR, three-sample SI-GFR, SS-GFR and SO-GFR, and (ii) determine the sensitivity of GFR to errors arising from different measurements.
Patients And Methods:
This study expanded on previous models of GFR measurement error incorporating biological data from 786 patients and realistic measurement errors. GFR median absolute error and the coefficient of variation (CV) were calculated for each method. A sensitivity analysis was carried out for individual measurement errors.
Results:
The median absolute error ranged between 1.2 and 2.3 ml/min/1.73 m, lowest for SS-GFR (4 h) and highest for SO-GFR. At higher rates of clearance, CV was less than 5% for all methods. CV increased rapidly when GFR decreased below a threshold ranging between 34 and 56 ml/min/1.73 m, lowest for three-point SI-GFR and highest for SO-GFR. SI-GFR and SS-GFR are most sensitive to injected activity quantification, but less sensitive to other measurement errors.
Conclusion:
Measurement errors are probably insignificant relative to biological variation for GFR of more than 60 ml/min/1.73 m, but become significant irrespective of biological variation at lower GFR, particularly in serial studies when GFR less than 25 ml/min/1.73 m. Limits of precision recommended in the 2018 British Nuclear Medicine Society guideline are appropriate for once-off GFR measurement, whereas slightly more stringent limits are proposed for serial studies.
Related Concept Videos
Glomerular Filtration Rate and its Regulation
GFR regulation involves two primary intrinsic controls: the myogenic and tubuloglomerular feedback mechanisms.
The myogenic...
Glomerular Filtration
Components of the Filtration Membrane
The filtration process involves three key layers: the glomerular endothelial cells, the basement membrane, and the podocyte-formed filtration slits.
Glomerular Filtration: Net Filtration Pressure
GBHP, with an average value of 55 mmHg, promotes filtration by pushing water and solutes through the filtration membrane. This is balanced by two opposing forces: CHP, a "back pressure" exerted against the filtration membrane by fluid already in the capsular space and renal...
Rate-Determining Steps
In a multistep reaction mechanism, one of the elementary steps progresses significantly slower than the others. This slowest step is called the rate-limiting step (or rate-determining step). A reaction cannot proceed faster than its slowest step, and hence, the rate-determining step limits the overall reaction rate.
The concept of rate-determining step can be understood from the analogy of a 4-lane freeway with a short-stretch of traffic-bottleneck caused due to...
Propagation of Uncertainty from Systematic Error
Propagation of Uncertainty from Random Error