Anodal right ventricular capture during left ventricular stimulation in CRT-implantable cardioverter defibrillators
Bernard Thibault1, Denis Roy, Peter G Guerra
1Department of Medicine, Montreal Heart Institute, Montreal, Quebec, Canada. ablation2000@bellnet.ca
Background:
Cardiac resynchronization therapy (CRT) has been shown to improve symptoms of patients with moderate to severe heart failure. Optimal CRT involves biventricular or left ventricular (LV) stimulation alone, atrio-ventricular (AV) delay optimization, and possibly interventricular timing adjustment. Recently, anodal capture of the right ventricle (RV) has been described for patients with CRT-pacemakers. It is unknown whether the same phenomenon exists in CRT systems associated with defibrillators (CRT-ICD). The RV leads used in these systems are different from pacemaker leads: they have a larger diameter and shocking coils, which may affect the occurrence of anodal capture.
Methods:
We looked for anodal RV capture during LV stimulation in 11 consecutive patients who received a CRT-ICD system with RV leads with a true bipolar design. Fifteen patients who had RV leads with an integrated design were used as controls. Anodal RV and LV thresholds were determined at pulse width (pw) durations of 0.2, 0.5, and 1.0 ms.
Results:
RV anodal capture during LV pacing was found in 11/11 patients at some output with true bipolar RV leads versus 0/15 patients with RV leads with an integrated bipolar design. Anodal RV capture threshold was more affected by changes in pw duration than LV capture threshold. In CRT-ICD systems, RV leads with a true bipolar design with the proximal ring also used as the anode for LV pacing are associated with a high incidence of anodal RV capture during LV pacing. This may affect the clinical response to alternative resynchronization methods using single LV stimulation or interventricular delay programming.
More Related Videos
Related Concept Videos
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Electrophysiology of Normal Cardiac Rhythm