HbA1c: a comparison of NGSP with IFCC transformed values
Gurdeep Singh Dhatt1, Mukesh M Agarwal, Bassam Bishawi
1Department of Pathology, Tawam Hospital, P.O. Box 15258, Al Ain, Abu Dhabi, United Arab Emirates. gurdeep1@emirates.net.ae
Summary
The International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) method for measuring Hemoglobin A1c (HbA1c) shows significant differences compared to the National Glycohemoglobin Standardization Program (NGSP) method. These differences mean results are not interchangeable, necessitating new diabetes management targets.
Area of Science:
- Clinical Chemistry
- Laboratory Medicine
- Diabetes Mellitus Diagnostics
Background:
- An International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) reference method for Hemoglobin A1c (HbA1c) is now available.
- This method correlates well with National Glycohemoglobin Standardization Program (NGSP) values.
- The study aimed to assess agreement between NGSP-certified and IFCC-calibrated methods for HbA1c measurement.
Purpose of the Study:
- To determine the agreement between NGSP-certified and IFCC-calibrated immunochemical methods for HbA1c.
- To evaluate differences in diabetes mellitus (DM) control classification for individual patients using both methods.
Main Methods:
- HbA1c was measured on 92 patient hemolysates using both NGSP and IFCC methods.
- Analysis included kappa statistic, linear regression, and McNemar's chi(2) test.
- The IFCC results were converted to NGSP values for comparison.
Main Results:
- Both methods demonstrated acceptable analytical performance with a kappa statistic of 0.65 and r(2)=0.937.
- A significant difference (p<0.0005) was observed in diabetes control classification for 22.8% of patients.
- The IFCC method, converted to NGSP values, differed significantly from the NGSP-certified method.
Conclusions:
- The IFCC and NGSP methods for HbA1c measurement yield non-interchangeable results.
- The observed differences are substantial enough to impact clinical decisions.
- The scientifically superior IFCC method should replace the NGSP method, requiring new HbA1c targets for diabetes management.