Effect of acquisition technique on radiation dose and image quality in multidetector row computed tomography coronary
Thomas C Gerber1, Brian P Stratmann, Ronald S Kuzo
1Department of Diagnostic Radiology, Mayo Clinic, Jacksonville, FL 32224, USA. gerber.thomas@mayo.edu
Rationale And Objectives:
We sought to examine effects of tube voltage and current on radiation dose and image quality for minimally invasive coronary angiography with a 16-slice multidetector row computed tomography (MDCT) scanner.
Materials And Methods:
We scanned the phantom used in the American College of Radiology Computed Tomography Accreditation Program at tube voltages of 80 and 120 kVp at 550, 650, and 750 mAseff, with and without a reduction in radiation dose by electrocardiographically (ECG) controlled tube current modulation (ECG pulsing).
Results:
Without ECG pulsing, the effective dose was 3 to 13 mSv. On average, a 50% increase in tube voltage led to increased radiation dose (215%), contrast-to-noise ratio (150%), and decreased image noise (-48%). On average, a 17% increase in mAseff led to increased radiation dose (17%) and contrast-to-noise ratio (4%) and decreased image noise (-9%). Dose reduction by ECG pulsing (simulated heart rate, 70 beats per minute) was 28%. With ECG pulsing, noise in images reconstructed during ventricular systole was double that in images reconstructed during ventricular diastole.
Conclusions:
These quantitative findings about the relationships among scan acquisition parameters, radiation dose, and image quality have practical implications for using ECG pulsing to reduce radiation doses in MDCT coronary angiography.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
X-ray Imaging
Imaging Studies for Cardiovascular System IV: CMRI