Performance of polymethyl pentene oxygenators for neonatal extracorporeal membrane oxygenation: a comparison with
Espeed Khoshbin1, Claire Westrope, Suneel Pooboni
1Division of Cardiac Surgery and Heart Link ECMO Centre, Glenfield University Hospital, Leicester, UK. khoshbinuk@yahoo.co.uk
Objective:
To review the performance of polymethyl pentene versus silicone oxygenators in terms of efficiency in priming and oxygenation, oxygenator resistance, requirements for coagulation proteins and consumption of blood products, for neonatal extracorporeal membrane oxygenation (ECMO) patients.
Study Design:
Forty consecutive neonates were selected retrospectively pre- and post-introduction of the new polymethyl pentene (PMP) oxygenators. They formed two equal groups. After calculation of the sample size, data were collected from ELSO registry forms and patient records. Results were analysed using parametric and non-parametric tests.
Results:
Neonatal PMP (N-PMP) oxygenators were smaller, faster and easier to prime. They were less efficient than silicone oxygenators, especially in carbon dioxide elimination, and, therefore, required higher sweeps. The preservation of coagulation proteins was significantly better, but there was no reduction in the consumption of blood products, despite having less than half the surface area and significantly lower blood path resistance.
Conclusion:
Small PMP oxygenators (Medos Hilite 800 LT) provide adequate gas exchange and offer technical advantages in terms of more efficient priming, reduced haemodynamic resistance and better control and preservation of coagulation proteins than silicone oxygenators.
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...