Clinical comparison of 6 aberrometers. Part 1: Technical specifications
Jos J Rozema1, Dirk E M Van Dyck, Marie-José Tassignon
1Vision Lab, University of Antwerp, Antwerp, Belgium. jos.rozema@ua.ac.be
Purpose:
To provide a detailed assessment of the techniques, technical features, and practical use of 6 aberrometers made available to our institution from September 2002 to January 2004.
Setting:
Department of Ophthalmology, University Hospital Antwerp, Antwerp, Belgium.
Methods:
A number of technical and practical parameters are listed for the Visual Function Analyzer (Tracey), the OPD-scan (ARK 10000; Nidek), the Zywave (Bausch & Lomb), the WASCA (Carl Zeiss Meditec), the MultiSpot Hartmann-Shack device, and the Allegretto Wave Analyzer including working principles, data acquisition, aberrometer alignment, wavefront calculation, and data analysis. Operator and patient comfort as well as practical advantages and disadvantages are discussed.
Conclusion:
All devices met at least half the following parameters: alignment, correction for source wavelength, data averaging, measurement quality check, and inhibition of accommodation.
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Uncertainty in Measurement: Reading Instruments
Influence of Earth's Curvature and Atmospheric Refraction on Leveling
Design Example: Measuring Distance Between Two Points with Obstructions
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer
Introduction and Methods of Leveling