Related Experiment Video
Updated: Aug 16, 2026

Sample Preparation for Computed Tomography-based Three-dimensional Visualization of Murine Hind-limb Vessels
Published on: October 7, 2021
Detection of small vessels with electron beam computed tomographic angiography using 1.5 and 3 mm collimator
Songshou Mao1, Jerold S Shinbane, Ronald J Oudiz
1Division of Cardiology, Harbor-UCLA Research and Education Institute, Torrance, CA 90502, USA.
Objectives:
To evaluate the effect of scanner collimation on the ability to detect small cardiac vessels using electron beam CT coronary angiography (EBA).
Materials And Methods:
EBA scans from 40 patients who underwent study on two separate occasions with 3 mm (initial scan) and 1.5 mm (follow-up scan) collimation protocols were analyzed. Vessels of <2 mm in diameter were identified.
Results:
The 1.5 mm collimation allowed 3-D visualization of 129 vessels<2 mm in diameter, while 3 mm collimation only allowed visualization of 89 vessels (p<0.001). The right coronary artery branches and distal LAD segments though were not displayed satisfactorily in almost half of the 3-D studies with either protocol.
Conclusions:
There was significant improvement in detection of small cardiac vessels with a 1.5 mm collimation EBA protocol compared to a 3 mm protocol. Both protocols though were insufficient for reliable visualization of the right coronary artery branches and distal LAD segments.