Evaluation of a Novel Laser-assisted Coronary Anastomotic Connector - the Trinity Clip - in a Porcine Off-pump Bypass Model
Published on: November 24, 2014
Noninvasive Doppler ultrasound evaluation of aortocoronary bypass
Insights
Ultrasound assessment of aortocoronary bypass grafts is a safe, inexpensive, and convenient screening tool. This method effectively evaluates graft patency in patients long after surgery.
Area of Science:
- Cardiovascular Surgery
- Medical Imaging
- Diagnostic Ultrasound
Background:
- Aortocoronary bypass grafting is a common surgical procedure.
- Long-term monitoring of graft patency is crucial for patient outcomes.
- Non-invasive methods for assessing graft patency are highly desirable.
Purpose of the Study:
- To evaluate the efficacy of ultrasound using a bidirectional Doppler's flowmeter for determining aortocoronary bypass graft patency.
- To assess the accuracy, reproducibility, and practicality of this ultrasound method for postoperative graft surveillance.
Main Methods:
- A prospective study involving 297 patients with 330 aortocoronary bypasses, assessed a mean of 8 months post-operation.
- Utilized a bidirectional Doppler's flowmeter for ultrasound assessment of graft patency.
- Compared ultrasound findings with angiographic results in a subset of 89 patients.
Main Results:
- Overall graft patency was 84.5%.
- Ultrasound demonstrated a specificity of 90% and sensitivity of 77% when compared to angiography.
- The method showed good reproducibility, was inexpensive, not time-consuming, safe, and convenient for patients.
Conclusions:
- Ultrasound with a bidirectional Doppler's flowmeter is a reliable and effective method for assessing aortocoronary bypass graft patency.
- The technique is suitable as a screening test for long-term follow-up of bypass grafts.
- Its safety, cost-effectiveness, and patient convenience make it a valuable tool in clinical practice.
Abstract:
Using ultrasound method, aortocoronary bypass patency was determined by means of bidirectional Doppler's flowmeter. In 8 (3%) of 302 subjects without aortocoronary bypass was recorded diastolic signal which might distort the results of postoperative evaluation of graft patency. Two-hundred and ninety-seven patients with 330 aortocoronary bypasses were studied at the mean time of 8 months after operation. The graft patency was 84.5%. Comparison of 89 ultrasound findings with angiographic results established considerable specificity (90%) and sufficient sensitivity (77%). Reproducibility of this method was also good. The method is inexpensive and not time consuming. The advantages are also its safety and convenience for the patient. In the present level of development it can be used mainly as screening test at long-term follow-up of aortocoronary graft patency.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Aortic Regurgitation II: Clinical Features and Diagnostic Tests
Acute Coronary Syndrome III: Diagnostic Studies