Related Experiment Video
Updated: Aug 16, 2026

Extraction and Detection of Geosmin and 2-Methylisoborneol in Water and Fish using High-Capacity Sorptive Extraction Probes and GC-MS
Published on: July 3, 2025
Coupling of SPME with MCC/UV-IMS as a tool for rapid on-site detection of groundwater and surface water contamination
G Walendzik1, J I Baumbach, D Klockow
1ISAS-Institute for Analytical Sciences, Bunsen-Kirchhoff-Str. 11, 44139 Dortmund, Germany.
Abstract:
The combination of headspace solid phase microextraction (HS-SPME) [1] with fast chromatographic pre-separation by means of multi-capillary columns (MCC) coupled to an ion mobility spectrometer equipped with a 10.6 eV photoionisation source was applied to rapid on-site monitoring of ground and surface water contaminations. Under field conditions, water contaminants were directly detectable down to the upper microg/L range. The potential of the developed method is illustrated by measurements of BTEX, naphthalene, chlorinated alkenes and chlorinated benzenes in water. Practical results from investigations of contaminated groundwater at a former gasworks site and groundwater from the site of a metal processing plant, focusing on the key substances benzene, naphthalene and tetrachloroethene, demonstrate the feasibility of the system for field studies.