Related Experiment Video
Updated: Aug 9, 2026

23:33
The WATCHMAN Left Atrial Appendage Closure Device for Atrial Fibrillation
Published on: February 28, 2012
Filter no reflow during percutaneous coronary interventions using the Filterwire distal protection device
I Porto1, R P Choudhury, P Pillay
1John Radcliffe Hospital, Oxford, UK. i.porto@lycos.com
International Journal of Cardiology
|August 9, 2005
Summary
Filter No Reflow (FNR) is a common complication during percutaneous coronary interventions using the Filterwire device. This reversible condition, often linked to debris, requires conservative management by interventionalists.
Area of Science:
- Interventional Cardiology
- Cardiovascular Devices
- Embolic Protection
Background:
- Distal protection devices are crucial for preventing embolization during percutaneous coronary interventions (PCI).
- The Filterwire device has been associated with a phenomenon termed Filter No Reflow (FNR), characterized by reversible flow reduction.
- FNR may be misinterpreted as true no-reflow and could impact device efficacy.
Purpose of the Study:
- To analyze the incidence and predictors of Filter No Reflow (FNR) during PCI with the Filterwire device.
- To investigate the relationship between FNR, collected debris, and angiographic findings.
- To differentiate FNR from true no-reflow and macroembolism.
Main Methods:
- A cohort of 58 patients (60 lesions) at high risk for embolization undergoing PCI of saphenous vein grafts (SVG) or native arteries (NV) with the Filterwire device was analyzed.
- Qualitative and quantitative angiographic analysis was performed.
- The volume of collected debris was estimated using a photographic technique.
Main Results:
- Filter No Reflow (FNR) occurred in approximately one-third of cases and was associated with visible filling defects within the filter basket, suggesting macroembolism.
- FNR showed a significant correlation with a higher volume of collected debris (36.97 mm³ vs. 11.31 mm³).
- The presence of FNR was not predicted by abciximab use, high thrombotic burden, increased stent volume, or need for predilatation; correlation with debris was only clear in the macroembolization group.
Conclusions:
- Filter No Reflow (FNR) is a common, reversible angiographic complication associated with the Filterwire device during PCI.
- Interventionalists should be aware of FNR and manage reduced flow conservatively.
- Both mechanical filter obstruction and other mechanisms, potentially involving debris or platelet aggregates, contribute to FNR.