Related Experiment Video
Updated: Aug 16, 2026

Biotribological Testing and Analysis of Articular Cartilage Sliding against Metal for Implants
Published on: May 14, 2020
The clinical performance of metal-on-metal as an articulation surface in total hip replacement
1The Arthritis Institute, 501 E. Hardy Street, 3rd Floor, Los Angeles, CA 90301, USA. patriciajpaul@yahoo.com
Abstract:
The metal-on-metal articulations in total hip arthroplasty (THA) were widely used between 1960 and 1975. The McKee-Farrar and other first-generation prostheses failed at a high rate because impingement caused early component loosening. The problem of early component loosening was corrected by improved component design and better manufacturing quality. Second-generation metal-on-metal total hip replacements have experienced short and medium-term success as assessed by Harris Hip Scores and patient self-assessment. The combined annual linear wear of the metal-on-metal femoral head and acetabular insert is less than 10 mm and osteolysis has only rarely been observed in association with well-fixed metal-on-metal total hip replacements. Hypersensitivity is not a common cause of loosening with second-generation hip replacements and remains to be proven as a definitive diagnosis in unusual cases of unexplained pain. More than 40 years of use has demonstrated no increase in the incidence of renal failure or cancer in patients with metal-on-metal total hip replacements. The scientific evidence of the results using the metal-on-metal articulations would recommend its continued use in any patient who does not have compromised renal function.