Effective Analysis of Human Exposure Conditions with Body-worn Dosimeters in the 2.4 GHz Band
Published on: May 2, 2018
Abnormal responses of electronic pocket dosimeters caused by high frequency electromagnetic fields emitted from
Shizuhiko Deji1, Kunihide Nishizawa
1Graduate School of Environmental Studies, Nagoya University, Furo-cho, Chikusa-ku, Nagoya 464-8602, Japan.
Abstract:
High frequency electromagnetic fields emitted from digital cellular telephones (cell phones) occasionally cause abnormally high and erroneous indicated dose readings on electronic pocket dosimeters (EPDs). Electric field strength distribution around a cell phone transmitting 1.5 GHz band with a maximum power of 0.8 W was analyzed by using an isotropic probe with tri-axial dipole antennas. Five types of EPDs were exposed to the fields for 50 s under configurations relative to the cell phone. The electric field distribution expanded around the phone's antenna and had a maximum electric field strength of 36.5 +/- 0.3 V m(-1). The cell phone gave rise to erroneous indicated dose readings on four out of five EPDs. The maximum value of erroneous indicated dosage for 50 s reached 1,283 microSv, which was about 2.6% of the annual effective dose limit of 50 mSv. The electromagnetic susceptibility of the EPDs was higher in the sections where the semiconductor detectors or electric circuit boards were located. The distance required to prevent electromagnetic interference differed for each EPD and ranged from 2.0 to 21.0 cm from the cell phone. The electric and magnetic field immunity levels of the EPDs varied from 9.2 V m(-1) to greater than 37.6 V m(-1), and from 0.03 A m(-1) to greater than 0.51 A m(-1). The EPDs displayed erroneous dose readings during exposure but recovered their normal performance after the cell phone ceased transmitting. The electromagnetic immunity levels of the EPDs were either equal to or greater than the IEC-standard. The immunity levels should be enhanced greater than the IEC-standard from the standpoint of radiation protection. The simplest and most reliable measure to prevent potential malfunction is to prohibit the radiation workers from carrying cell phones to their workplace.
More Related Videos
06:21Diffuse Optical Spectroscopy for the Quantitative Assessment of Acute Ionizing Radiation Induced Skin Toxicity Using a Mouse Model
Published on: May 27, 2016
08:23An Automated Microscopic Scoring Method for the γ-H2AX Foci Assay in Human Peripheral Blood Lymphocytes
Published on: December 25, 2021
Related Concept Videos
Biological Effects of Radiation
Mutations
Chromosomal Alterations Are Large-Scale Mutations
While point mutations are changes in a single nucleotide in...
Electronic Distance Measuring Instruments