Abstract:
A method for the direct determination of trace lead in chestnut by Graphite Furnace Atomic Absorption Spectrometry (GFAAS) with slurry sampling was developed in the present paper. Ammonium ortho-phosphate was used as a matrix modifier. The effects of slurry preparation, particle size of sample, matrix modifier, ashing temperature, atomization temperature, and common coexisting components on the determination of lead were studied. Under the optimized operating conditions, the detection limit, the relative standard deviation (RSD), and the recoveries of standard addition for this method were 0.47 ng x mL(-1), 6.1% (n = 11) and 90%-106%, respectively.
Related Concept Videos
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Precipitation Gravimetry
In determining nickel by gravimetric analysis, a precipitant of ethanolic dimethylglyoxime is added to a hot nickel salt solution. This is quickly followed by the dropwise addition of dilute ammonia solution until precipitation occurs. A...