Related Experiment Video
Updated: Aug 16, 2026

Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Comparison of the ICare rebound tonometer with the Goldmann tonometer in a normal population
P Fernandes1, J A Díaz-Rey, A Queirós
1Department of Physics (Optometry), School of Sciences, University of Minho, Braga, Portugal. pfernandes@fisica.uminho.pt
Abstract:
The aim of this study was to evaluate the accuracy of measurement of intraocular pressure (IOP) using a new induction/impact rebound tonometer (ICare) in comparison with the Goldmann applanation tonometer (AT). The left eyes of 46 university students were assessed with the two tonometers, with induction tonometry being performed first. The ICare was handled by an optometrist and the Goldmann tonometer by an ophthalmologist. In this study, statistically significant differences were found when comparing the ICare rebound tonometer with applanation tonometry (AT) (p < 0.05). The mean difference between the two tonometers was 1.34 +/- 2.03 mmHg (mean +/- S.D.) and the 95% limits of agreement were +/-3.98 mmHg. A frequency distribution of the differences demonstrated that in more than 80% of cases the IOP readings differed by <3 mmHg between the ICare and the AT. In the present population the ICare overestimates the IOP value by 1.34 mmHg on average when compared with Goldmann tonometer. Nevertheless, the ICare tonometer may be helpful as a screening tool when Goldmann applanation tonometry is not applicable or not recommended, as it is able to estimate IOP within a range of +/-3.00 mmHg in more than 80% of the population.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Comparing Experimental Results: Student's t-Test
Behrens–Fisher Test
This test is...