Related Experiment Video
Updated: Aug 16, 2026

Application and Methodology of the Non-destructive 19F Time-domain NMR Technique to Measure the Content in Fluorine-containing Drug Products
Published on: August 22, 2017
Direct determination of niflumic acid in a pharmaceutical gel by ATR/FTIR spectroscopy and PLS calibration
C Boyer1, B Brégère, S Crouchet
1Laboratoire de Chimie Analytique, EA 2962, Université Victor Segalen Bordeaux 2, 146 rue Léo Saignat, 33076 Bordeaux Cedex, France.
Abstract:
A simple, rapid and convenient analytical method without sample handling procedure is proposed for the determination of niflumic acid in a pharmaceutical gel with attenuated total reflectance/Fourier transform infrared spectroscopy (ATR/FTIR). A partial least square (PLS) calibration model for the prediction of niflumic acid contents was developed using 81 and 27 spectra of standard gels as training and validation sets, respectively. The used spectral range of niflumic acid for the establishment of this model was 2300-1100 cm(-1). All spectra were obtained in the transmittance mode, then normalized and first derivative transformed. The model yielded a regression coefficient R2 equal to 1 for the training set and a root mean square error of prediction (RMSEP) equal to 0.2 for the validation set. The percentage recoveries of the method for the analysis of Niflugel ranged from 96.60 to 101.02%.